OWON HSA2300

OWON HSA2300 Series Handheld Multifunction Spectrum Analyzer, Oscilloscope, and Multimeter

Model: HSA2300 Series (HSA2307, HSA2310, HSA2320)

Brand: OWON

1. Entwodiksyon

The OWON HSA2300 Series is a versatile 3-in-1 handheld device designed for comprehensive electronic measurement and analysis. It integrates the functionalities of a Spectrum Analyzer, a Digital Oscilloscope, and a True RMS Multimeter into a single portable unit. This manual provides essential information for safe operation, setup, usage, maintenance, and troubleshooting of your device.

OWON HSA2300 Series 3-in-1 Multifunction Device Overview
Figure 1.1: OWON HSA2300 Series 3-in-1 Multifunction Device Overview. This image highlights the three core functions: 3GHz Spectrum Analyzer, 200MHz Oscilloscope, and 4 1/2 digit Multimeter.

2. Enfòmasyon sou Sekirite

Before using this product, please read the following safety information carefully to prevent personal injury and damage to the product or any other connected devices. This product must only be used within its specified range.

2.1. Regleman ak Senbòl Sekirite

Terms in this manual:

  • Avètisman: Endike kondisyon oswa pratik ki ka lakòz blesi oswa pèt lavi.
  • Atansyon: Indicates conditions or practices that could result in damage to this product or other property.

Terms on the product:

  • Danje: Endike yon danje imedya oswa posiblite aksidan.
  • Avètisman: Endike yon danje oswa blesi posib.
  • Atansyon: Endike potansyèl domaj nan enstriman an oswa lòt pwopriyete.

Safety Symbols on the product:

  • Danjere Voltage
  • Gade Manyèl
  • Tèminal Latè Pwoteksyon
  • Chasi Ground
  • Tès Tès

2.2. Kondisyon Sekirite Sosyal

  • Sèlman yon moun ki kalifye ta dwe fè antretyen entèn.
  • Caution: Check all Terminal Ratings. Pou evite danje dife oswa chòk, tcheke tout evalyasyon ak mak sou pwodui sa a. Gade manyèl itilizatè a pou plis enfòmasyon sou evalyasyon anvan ou konekte ak enstriman an.
  • Pa opere san kouvèti.
  • Evite sikwi ekspoze. Fè atansyon lè w ap travay sou sikwi ekspoze pou evite risk pou chòk elektrik oswa lòt blesi.
  • Do not operate if any damage. If you suspect damage, have it inspected by qualified service personnel.
  • Pa opere nan damp kondisyon oswa atmosfè eksplozif.
  • Kenbe sifas pwodwi pwòp epi sèk.
  • Sèvi ak ekipman ki pa an akò ak metòd manifakti a espesifye ka domaje pwoteksyon ekipman an bay.
  • Avètisman: To prevent electric shock or fire, use a suitable power adapter. Only power adapters dedicated to this product and approved for use in the country of use may be used.
  • Avètisman: The two channels of the oscilloscope are non-isolated. Ensure a common reference when measuring. Do not connect ground wires of two probes to non-isolated places with different DC electrical levels to avoid short circuits.
  • If the input port is connected to a voltage higher than 42V (30Vrms) or a circuit with a peak value of more than 4800 VA, use only voltage probes, test wires, and adapters with proper insulation specified by OWON.
  • Before use, check all probes and accessories for mechanical damage. Replace if damaged.
  • Remove all unused multimeter test probes, oscilloscope probes, and accessories (power adapter, USB, etc.) before use.
  • Plug the power adapter into the AC socket first, then connect it to the oscilloscope.
  • Lè tès nan yon anviwònman CAT II, ​​pa konekte yon voltage higher than 400 V to any input port or with a voltage diferans ki gen plis pase 400 V nan pò a opinyon izole.
  • Pa sèvi ak yon D' voltage higher than the rated value. Pay special attention when using 1:1 test wires, as the probe voltage pral dirèkteman transmèt nan osiloskop la.
  • Pa manyen metal vid BNC oswa ploge bannann lan.
  • Pa mete objè metal nan konektè a.
  • Sèvi ak osiloskop la sèlman nan fason espesifye a.

3. Pwodwi souview

3.1. Front Panel and Keys

Front Panel Layout of HSA2300
Figure 3.1: Front Panel Layout of the HSA2300. This image shows the LCD display, multi-functional key area, power button, Type-C charging port, multimeter input terminals, oscilloscope input terminals, and spectrum SMA input port.
  1. CH1 and CH2 input connectors (Oscilloscope).
  2. Spectrum measured signal input connector.
  3. Display area (3.5-inch LCD).
  4. F1 - F4 keys: Multi-function keys for menu selection.
  5. HOR key: Adjusts horizontal time base and displacement.
  6. Return key: Returns to the previous menu or closes the menu.
  7. Measurement menu key (oscilloscope), range key (multimeter), or reset key (spectrum).
  8. Zoom or move keys (directional arrows): For waveform movement, time base changes, cursor movements, and trigger level changes.
  9. Automatic setting key (oscilloscope), automatic range key (multimeter), or automatic search key (spectrum).
  10. Stop / run key (oscilloscope), value hold key (multimeter), or output/disable signal key (spectrum).
  11. Multimeter input terminals.
  12. Trigger menu key (oscilloscope), relative value key (multimeter), or select numeric key (spectrum).
  13. Kle switch pouvwa.
  14. Save settings key.
  15. System settings key.
  16. Switch key for oscilloscope and multimeter working state.
  17. CH1 / CH2 - channel switch key.

3.2. Side Panel

Side Panel Layout of HSA2300
Figure 3.2: Side Panel Layout of the HSA2300. This image shows the probe compensation output, USB Type-C charging/communication port, and the built-in bracket.
  1. Probe compensation: 3.3V/1kHz square wave signal output.
  2. Charging or USB communication interface (Type-C).
  3. Bracket for stable support.

3.3. Lis anbalaj

OWON HSA2300 Packing List
Figure 3.3: Packing List. The package includes the Main Unit, Oscilloscope Probe, Multimeter Leads, USB Type-C Cable, Soft Bag, and Quick Guide.

Pake estanda a gen ladan:

  • 1 x Main Unit (OWON HSA2300 Series)
  • 1 x USB kab
  • 1 set x Passive Probes (2 pieces)
  • 1 set x Multimeter Probes (one red and one black)
  • 1 x Soft Bag
  • 1 x Gid rapid

4. Enstalasyon ak Premye Enspeksyon

4.1. Enspeksyon jeneral

Upon receiving your new oscilloscope, perform the following checks:

  1. Check for transportation damage: Inspect the packaging carton and foam for serious damage. If found, do not discard packaging until the device and accessories pass electrical and mechanical tests.
  2. Tcheke Pwodwi pou Telefòn: Refer to the packing list (Section 3.3) to ensure all accessories are present and undamaged. Contact your distributor if any items are missing or damaged.
  3. Check the complete instrument: Inspect the instrument for any physical damage. If the instrument does not work normally or fails performance tests, contact your distributor. Keep the original packaging if transportation damage is suspected.

4.2. Konpansasyon Sond

When connecting a probe to any input channel for the first time, it is essential to adjust the probe compensation to match the input channel. Incorrect compensation can lead to measurement errors.

  1. Set the probe menu attenuation coefficient to 10X in the oscilloscope settings.
  2. Set the switch on the oscilloscope probe to 10X and connect it to the CH1 channel. Align the slot on the probe with the BNC connector, insert, and turn to tighten.
  3. Connect the probe tip and ground clamp to the probe compensator's signal output and reference ground terminals, respectively.
  4. Press the "Auto" key on the front panel. A square wave display (1kHz/3.3Vpp) should appear.
  5. Check the displayed waveform and adjust the probe until the compensation is correct. Repeat for CH2 if necessary.
True RMS Measurement vs Conventional Measurement
Figure 4.1: True RMS Multimeter. This image illustrates the difference between conventional measurement (accurate only with sine waves) and True RMS measurement (accurately measures all current waveforms).

4.3. Konekte ak PC

The oscilloscope supports communication with a PC via USB for data storage, analysis, display, and remote control. Download the Oscilloscope software package from the official OWON website and follow the installation prompts.

  1. Koneksyon: Use a USB data cable to connect the USB Device port on the right panel of the Oscilloscope to a USB port on your PC.
  2. USB Port Settings: On the oscilloscope, navigate to SistèmF4USB, and switch the USB protocol type to HID (Human Interface Device).
  3. Run the Oscilloscope software. The connection information in the bottom right corner of the software should turn green, indicating a successful connection.

5. Operasyon Enstriksyon

5.1. Mòd Osiloskop

To use the oscilloscope function, ensure the device is in oscilloscope mode (use the switch key for working state).

5.1.1. Entèfas itilizatè

The oscilloscope interface displays waveform data and various status indicators:

Oscilloscope Interface with One-Click Conversion
Figure 5.1: Oscilloscope Interface with One-Click Conversion. This image demonstrates the ability to switch between different measurement views (Voltage, Current, Spectrum) with precise measurement.
  • Trigger Status: Indicates Auto, Trig, Ready, Scan, or Stop modes.
  • Kouri/sispann: Indicates if data acquisition is active or paused.
  • Time Base Display: Shows the current horizontal time scale.
  • Trigger Horizontal Position: Pointer indicating the trigger point.
  • Enfòmasyon sou chanèl la: Voltage position and coupling mode (DC, AC, Ground).
  • Zòn Ekspozisyon Fòm Vag: Where the signal waveforms are shown.

5.1.2. Functional Check

To quickly verify proper operation:

  1. Press the power switch key. The device will perform self-checks and display the startup screen.
  2. Set the oscilloscope probe switch to 10X and connect to CH1.
  3. Connect the probe tip and ground clamp to the probe compensator.
  4. Press the "Auto" key. A 1kHz/3.3Vpp square wave should be displayed.

5.1.3. Vertical System

Adjust vertical scale, position, and other channel settings. Each channel has independent settings.

  • Pozisyon vètikal: Sèvi ak CH1/CH2 keys and directional arrows to move the waveform up/down. Press both arrows simultaneously to center.
  • Vertical Volt/Div Setting: Sèvi ak CH1/CH2 keys and directional arrows to change the volts/div range (10mV/div-10V/div for 1X probe).
Vertical System Settings
Fonksyon Menu Anviwònman Deskripsyon
Chanje ON/OFF Turn on/off waveform display.
Couplage DC, AC, GND Normal sampling (DC/AC components), block DC component, disconnect input.
Sonde 1X, 10X, 100X, 1000X, 10000X Choose value based on probe attenuation factor for accurate reading.
Bandwidth 20MHz, Full band Limit bandwidth to 20MHz or use full bandwidth.

5.1.4. Horizontal System

Peze a HOR key to enter the horizontal system setting menu. Use directional keys to change the horizontal scale (time base) and horizontal trigger position. Press both arrows simultaneously to center the horizontal position.

Anviwònman sistèm orizontal
Fonksyon Menu Anviwònman Deskripsyon
Acqu Mode Sample, Peak Detect Normal sampling or detect interference glitches.
Longè 4K, 8K Select length to be recorded.
XY mòd ON/OFF Turn XY mode on or off.
Rafrechi Segondè, ba Set refresh rate.
Hor center - Set horizontal trigger position to middle of screen.
Counter ON/OFF Turn on/off counter.

5.1.5. Automatic Measurement

Peze Range epi F1 to perform automatic measurements. Up to 9 types of measurements can be displayed. Available measurements include Frequency, Period, Amplitude, Maximum, Minimum, Peak-to-Peak value, Mean, RMS, Rise Time, Fall Time, +PulseWidth, -PulseWidth (depending on bandwidth model).

Oscilloscope Measurement Features
Figure 5.2: Oscilloscope Measurement Features. This image illustrates two acquisition modes (Normal sampling, Peak sampling), 12 auto-measurements (Period, Frequency, Amplitude, etc.), cursor measurement, and save function (4 screenshots, 4 CSV waveform datasets).

5.1.6. Cursor Measurement

Peze Range epi F2 to activate cursor measurement. You can select CH1 Voltage, CH2 Voltage, or Time cursors. Use directional keys to move cursors A and B. Link A and B to move both simultaneously.

5.1.7. Trigger System

The trigger determines when the oscilloscope acquires and displays data. The HSA2300 series uses edge trigger mode. Trigger settings are displayed at the bottom right of the screen (e.g., CH1:DCF-20.0mV).

Trigger System Settings
Fonksyon Menu Anviwònman Deskripsyon
Sous CH1, CH2 Set Channel 1 or 2 as the source trigger signal.
Couplage AC, DC Prevent DC component or allow all components to pass.
Kalite Oto, nòmal, sèl Waveforms acquired without trigger, only when trigger conditions met, or sampled and stopped on trigger.
Pant Kwen k ap monte, Kwen k ap tonbe Trigger on rising or falling edge.
Trigger center - Trigger position set in the middle of the waveform.
Fòs - Forcibly generates a trigger signal (used in normal/single modes).

Trigger electrical level: Peze Trig/△ and directional keys to adjust the amplitude level the signal must cross.

5.1.8. Save Settings

Peze a Sove key to store oscilloscope settings, reference waveforms, and files.

  • Configured: Save/recall current parameter settings.
  • Reference Waveform: Compare actual waveform with a stored reference.
  • File: Save waveforms (CSV) or screen images (BMP). Files can be accessed via USB (MSC mode).

5.1.9. Anviwònman Sistèm

Peze a Sistèm key to access system functions.

  • Ekspozisyon: Adjust brightness, backlight time, menu display time.
  • Sistèm: Set language, automatic shutdown time, view instrument info (model, serial, version), upgrade firmware.
  • Anviwònman defo: Reyajiste nan défaut faktori.
  • Koneksyon USB: Select HID (Human Interface Device) for PC control or MSC (Mass Storage Class) for file transfè.
  • Anviwònman faktori: Reyajiste tout paramèt yo ak parèy faktori yo.

5.1.10. Otomatik Kalibrasyon

The automatic correction program optimizes the oscilloscope for accurate measurements. Execute this program if the ambient temperature changes by more than 5°C.

To perform: Disconnect all probes/wires. Press SistèmF4F3 ("Auto calibration"). Confirm readiness and proceed.

5.2. Mòd Miltimèt

The multimeter function uses four 4-mm safety banana plug input ends: A, mA, COM, and V→C.

Entèfas Miltimèt
Figure 5.3: Multimeter Interface. This image shows the display for voltage measurement, indicating measurement type, range, hold function, and measurement value.

Kalite mezi:

  • ⎯DCV: DC voltage mezi
  • ∼ACV: AC voltage mezi
  • ⎯DCA: DC mezi aktyèl la
  • ∼ACA: AC aktyèl mezi
  • Ω Resist: Mezi rezistans
  • ‡ Diode: Mezi dyòd
  • ♫ Cont: On/Off measurement (Continuity)
  • ⋃ Cap: Mezi kapasite

Range Indication: "Manual" for manual range, "Auto" for automatic range.

Kenbe: Keeps the current reading on the display.

Relative Value Measurement: Available for DC current, DC voltage, ak mezi rezistans.

5.3. Spectrum Analyzer Mode

Peze a Mode key to switch to the spectrum interface.

5.3.1. Spectrum Interface

Spectrum Analyzer Interface
Figure 5.4: Spectrum Analyzer Interface. This image shows the spectrum display with reference level, attenuator, scale/div, bandwidth, USB connection indicator, battery power indicator, and marker.

Key elements of the spectrum interface:

  • Nivo referans
  • Atenuateur
  • Scale/Div
  • Bandwidth
  • USB cable inserted indicator
  • Endikasyon pouvwa batri
  • Makè
  • Zòn ekspozisyon Waveform
  • Stop frequency, Center frequency, Start frequency
  • Marker parameter
  • Ampechèl latitid

5.3.2. Parameter Input Interface

Parameters are modified using the numeric keypad, Trig/△ button, and directional keys.

  • X (cancel button): Clears all input characters.
  • <- (back button): Deletes characters from right to left.
  • Unit buttons: Select appropriate units (X, GHz, MHz, kHz, Hz).
  • Antre bouton: Completes parameter input.
  • Numeric keys (0-9): For inputting values.

5.3.3. Frequency System

Access the frequency menu via F1 (Spectrum menu → Frequency) or directly via CH1/2.

  • Start Frequency: Activates the start frequency setting. Note that span and center frequency change automatically.
  • Stop Frequency: Activates the stop frequency setting.
  • Frekans Sant: Activates the center frequency setting.
  • span: Sets the sweep width and frequency range. Options include Full Span (maximum sweep width) and Zero Span (sweep width to 0, horizontal axis represents time).

5.3.4. Amplitude System

Aksede a amplitude menu via F1 (Spectrum menu → Amplitude) or directly via HOR.

  • Nivo Referans: Sets the maximum power or voltagvalè ki parèt.
  • Scale/Div: Sets the size of each vertical grid unit (1, 2, 5, or 10 dB per grid).
  • Atenuasyon: Adjusts the RF front-end attenuator to ensure large signals pass with low distortion.
  • Reference Unit: Chwazi la amplitude unit ([dBm], [dBmV], [dBµV], [W], [V]).
  • Reference Offset: Adds an offset value to compensate for gain or loss between the device under test and the spectrum analyzer input.

5.3.5. Bandwidth System

Access the bandwidth menu via F1 (Spectrum menu → Bandwidth).

  • RBW (Resolution Bandwidth): Selects the resolution bandwidth. Reducing RBW increases frequency resolution but may lengthen sweep time.
  • VBW (Video Bandwidth): Selects the video bandwidth. Reducing VBW smooths the spectrum line and differentiates small signals from noise, but may lengthen sweep time.

5.3.6. Marker System

Access the marker menu via F1 (Spectrum menu → Marker).

  • Makè: Selects frequency marker type (Normal, Delta, All). Normal measures X/Y values at a point. Delta measures delta values between two points.
  • Pik: Finds maximum peak, peak-to-peak, next peak, left/right peak, and sets peak search mode/height.

5.3.7. Trigger System (Spectrum)

Access the trigger menu via F1 (Spectrum menu → Trigger).

  • oto: Toggles auto trigger on/off.
  • Videyo: Toggles video trigger on/off.
  • pre-amplifier: Enables/disables the RF front-end amplifier for weak signals.

5.3.8. Automatic Research

Peze a Oto key to activate automatic search. This scans the frequency range, adjusts parameters to optimal settings, and allows for one-touch signal search.

6. Antretyen ak netwayaj

6.1. Antretyen jeneral

  • Do not store the instrument in direct sunlight for prolonged periods.
  • Atansyon: Do not allow spray, liquid, or solvent to touch the instrument or probes.

6.2. Netwayaj

Regularly check and clean the instrument and probes:

  1. Wipe external surfaces with a soft cloth to remove dust. For the LCD, be careful not to scratch the protection screen.
  2. Siye enstriman an ak anonsamp (not dripping) soft cloth. Disconnect power supply first. A mild detergent or water can be used.
  3. Do not use abrasive chemical cleaning agents to avoid damage.

Avètisman: Ensure the instrument is completely dry before re-energizing to prevent electrical short circuits or personal injury.

6.3. Battery Charging and Replacement

The device uses a replaceable 18650 Lithium Battery (2600mAh*2, 3.7V).

Chaje:

  • If the device does not turn on after long-term storage, pre-charge for 0.5 to 1 hour using the attached adapter.
  • Charge regularly if not used for a long time to prevent over-discharge.
  • Charging via power adapter: Connect the oscilloscope to the power socket using the USB Type-C cable and power adapter.
  • Charging via USB interface: Connect the oscilloscope to a computer or other equipment via USB. Ensure the power supply equipment has sufficient load capacity.
Battery Charging/Discharging Times
Modèl Chaje Dechaje
For models <100MHz ≥4.5h ≥4h
For models ≥100MHz ≥4h ≥3h

Endikatè batri:

  • : Power-on charging status.
  • 🔋: Battery power supply.
  • 🔋 (low): Approximately five minutes of use time left. Charge immediately.

To avoid overheating during charging, the ambient temperature must not exceed the allowable value in the technical specifications.

Ranplasman: The battery generally does not need replacement. If necessary, it must be replaced by qualified personnel using only lithium batteries of the same specification.

NAN. Depanaj

  1. Osiloskop la pa ka limen.

    The battery may be completely exhausted. Charge the battery for 15 minutes without turning on the oscilloscope, then try again. If it still doesn't turn on, contact support.

  2. Osiloskop la etenn apre kèk segonn nan demaraj.

    The battery may be exhausted. Check the battery symbol; if it indicates low, charge the device.

  3. After switching to the multimeter, the measurement type is displayed as 'E'.

    The measurement type may not be selected. Press F4 to select the corresponding measurement type. If 'E' persists, restart the oscilloscope.

  4. In oscilloscope state, the measured voltage amplitude is 10 times larger or smaller than the actual value.

    Check if the channel attenuation coefficient is consistent with the actual probe error ratio.

  5. In oscilloscope state, there is a waveform display, but it cannot be stabilized.
    • Check if the source item in the trigger mode menu matches the signal channel used.
    • Check if the trigger electrical level has exceeded the waveform range. Adjust parameters to stabilize the waveform.
  6. In oscilloscope state, nothing is displayed after pressing RUN/STOP.

    Check if the trigger mode is normal or single, and if the trigger electrical level is out of range. Center the trigger electrical level or set the trigger mode to automatic. Press Oto to automatically adjust settings.

  7. In oscilloscope state, the display speed becomes slower when average value sampling is set or duration is longer.

    Sa a se konpòtman nòmal.

8. Espesifikasyon teknik

All technical specifications apply to probes with attenuation switch set to 10X. The instrument must operate continuously for more than 30 minutes at the specified operating temperature. If the temperature variation exceeds 5°C, perform automatic correction via System Settings.

8.1. Oscilloscope Specifications

Karakteristik Deskripsyon
Rezolisyon vètikal8 bit
Chèn2
Sampmetòd lingSampling, peak detection
An tan reyèl samppousantaj lingHSA2105: 125 MSa/s (Dual Channel)
HSA2307: 250 MSa/s (Single Channel)
HSA2310: 500 MSa/s
HSA2320: 1 GSa/s
Waveform refresh rate10,000 wfms/s
Antre koupleDC, AC, ground
Enpedans Antre1 MΩ±2%, in parallel with 16 pF±10 pF
Sonde atténuation1X, 10X, 100X, 1000X, 10000X
D 'maksimòm voltage400 V (DC + AC, PK - PK)
Limite band20 MHz, Full bandwidth
Sampranje pousantaj ling0.25 Sa/s - 250 MSa/s
Entèpolasyon fòm ond(Senx)/x
Sweep speed range (S/div)HSA2105/2307: 5ns/div - 1000s/div
HSA2310/2320: 2ns/div - 1000s/div (Stepping in 1-2-5 way)
Tan baz presizyon± 100 ppm
Longè dosye8K oswa 4K si ou vle
Sensitivity (Volt/div) range10 mV/div - 10 V/div
Ranje deplasman±6 div
±2 V (10 mV/div - 200 mV/div); ±100 V (500 mV/div - 10V/div)
Pleasant analogHSA2105: 50 MHz
HSA2307: 70 MHz
HSA2310: 100 MHz
HSA2320: 200 MHz
Single bandwidthPleasant konplè
Low frequency response (AC coupling, -3dB)≥10 Hz
Rise time (typical on BNC)HSA2105: ≤ 5 ns
HSA2307: ≤ 3.5 ns
HSA2310: ≤ 1.75 ns
HSA2320: ≤ 1.75 ns
DC jwenn presizyon3%
KurseurΔV, ΔT
Otomatik MeziPeriod, Frequency, Mean, PK-PK, Max, Min, Amplitude, RMS, Rise Time, Fall Time, +PulseWidth, -PulseWidth
Deklanche SousCH1, CH2
Kalite deklancheEdge
Deklanche koupleDC, AC
Kalite deklancheOto, nòmal, sèl
Trigger electrical level range±4 div from the center of the screen
Trigger electrical level accuracy±0.3 div
Deklanche deplasmanAccording to Record length and time base
Edge triggering slopeKwen k ap monte, kwen k ap tonbe
Probe compensator output voltage (tipik)3.3Vpp, High-Z
Probe compensator frequency (typical)Square wave 1 kHz (±1%)

8.2. Multimeter Specifications

Karakteristik Deskripsyon
Ekspozisyon dijital20,000 lekti
Kalite meziVoltage, current, resistance, capacitance, on/off, diode
Maksimòm Antre voltageAC: 750V DC: 1000V
Maksimòm Antre aktyèlAC: 10A DC: 10A
Fonksyon debaz Range Minimòm rezolisyon Presizyon
DC voltage20.000mV - 1000.0V0.001mV - 0.1V±(0.3%+5dig) to ±(0.5%+15dig)
AC voltage20.000mV - 200.00V0.001mV - 0.01V± (0.8% + 10 fouye)
DC aktyèl20.000mA - 10.000A0.001mA - 1mA±(1.0%+10dig) to ±(2.5%+10dig)
AC aktyèl20.000mA - 10.000A0.001mA - 1mA±(1.0%+10dig) to ±(2.8%+10dig)
Rezistans200.00Ω - 100.00MΩ0.01Ω - 0.01MΩ±(0.8%+3dig) to ±(5.0%+10dig)
Kapasite2.000nF - 2.000mF1pF - 1uF± (3.0% + 10 fouye)
On/Off test√ (<50Ω)--
Dyòd tès√ (<0-2V)--
Ranje oto--
TRMS--

8.3. Spectrum Analyzer Specifications

Karakteristik Deskripsyon
Gamme frekansHSA2105: 9 kHz to 1.5 GHz
HSA2307/2310/2320: 9 kHz to 3.0 GHz
Rezolisyon Frekans100 Hz
Frequency Span RangeZero span, 300 Hz to maximum bandwidth
Resolution Bandwidth0.2, 1, 3, 10, 30, 100, 300, 600, 850 kHz
Max. Input Level (Continuous power)Attenuator= 30 dB: +20 dBm (100 mW)
Max. damage level+20 dBm (100 mW)
Frekans Referans Entèn26MHz
Display Average Noise Level (DANL)10MHz to 1 GHz: <-120 dBm
1 GHz to 2 GHz: <-120 dBm
2 GHz to 3 GHz: <-115 dBm (Preamp Off)
10MHz to 1 GHz: <-140 dBm
1 GHz to 2 GHz: <-140 dBm
2 GHz to 3 GHz: <-135 dBm (Preamp Sou)
Level display Logarithmic level0.01 dB a 99 dB
Level display Linear level0 to reference level
Number of display points301
Kantite tras1
Inite yodBm, dBmV, dBµV, W, V
Input attenuation setting range0 dB to 30 dB, step by 1 dB
pre-amppi vivan20 dB (typical value)
Enpedans Antre RF50 Ω, typical
RF Antre ConnectorSMA fi konektè

8.4. Espesifikasyon teknik jeneral

Karakteristik Deskripsyon
Kalite ekspozisyon3.5-pous koulè LCD ekspozisyon
Rezolisyon ekspozisyon320 orizontal × 240 piksèl vètikal
Montre koulè65536 koulè
Montre kontrasReglabl
Ekipman pou pouvwaDC INPUT: 5VDC, 2A
Konsomasyon pouvwaHSA2105/2307/2310: <5 W
HSA2320: <6 W
Batri2600mAh*2 (3.7V, 18650)
Tanperati travay0℃ - 40℃
Tanperati depo-20℃ - +60℃
Imidite relatif≤90%
Wotè opere3,000 mèt
Non-operating height15,000 mèt
Metòd refwadismanRefwadisman natirèl
Dimensions (L×H×D)198 mm × 96 mm × 38 mm
Weight (main unit, without battery)Apeprè 0.6 kg
Entèval kalibrasyonRecommended: one year
OWON HSA2300 Dimensions
Figure 8.1: OWON HSA2300 Dimensions. The device measures 96mm (W) x 198mm (H) x 38mm (D) and weighs approximately 0.6 kg (main unit).

9. Konsèy itilizatè yo

  • For optimal performance, ensure probes are properly compensated before critical measurements.
  • Regularly check battery status, especially before extended use, to avoid unexpected shutdowns.
  • When connecting to a PC, remember to set the USB mode to HID for control or MSC for file transfè.
  • If experiencing unstable waveforms, verify trigger settings and probe attenuation match your signal.

10. Garanti ak sipò

10.1. General Warranty

OWON warrants that the product will be free from defects in materials and workmanship for a period of 3 years from the date of purchase by the original purchaser. Accessories such as probes and adapters have a 12-month warranty. This warranty is non-transferable.

If the product is defective during the warranty period, OWON will repair or replace it. Replaced parts or products may be new or reconditioned. All replaced items become the property of OWON.

To obtain service, notify OWON before the warranty expires. The customer is responsible for packaging and shipping the defective product to the service center, along with proof of purchase.

This warranty does not cover defects or damage caused by improper use, inadequate maintenance, unauthorized repairs, improper connection to incompatible equipment, use of non-OWON supplies, or modifications that increase servicing difficulty.

Excepting the after-sales services provided, OWON offers no other guarantees, express or implied, including for marketability or special-purpose acceptability. OWON is not responsible for any indirect, special, or consequential damages.

10.2. Sipò pou pwodwi

For product support, visit the official OWON websit: www.owon.com.hk/download

You can also refer to the comprehensive user manual PDF available at: OWON HSA2000 Dual Channel Series Handheld Spectrum User Manual (PDF)

Enfòmasyon pou kontakte:
Fujian LILLIPUT Optoelectronics Teknoloji co, Ltd.
No. 19, Heming Road, Lantian Industrial Zone, Zhangzhou 363005 P.R. China
Tel: +86-596-2130430
Faks: +86-596-2109272
Imèl: info@owon.com.cn
Web: www.owon.com.cn

Dokiman ki gen rapò - HSA2300

Preview Manyèl Itilizatè Osiloskop Pòtab OWON HDS200 Seri - Operasyon ak Espesifikasyon
Manyèl itilizatè konplè sa a bay enstriksyon detaye pou itilize osiloskop pòtab OWON HDS200 Seri Doub Chanèl yo, ki gen ladan modèl HDS272(S) ak HDS242(S). Li kouvri enfòmasyon sou sekirite, enspeksyon jeneral, osiloskop, miltimèt, ak fonksyon jeneratè fòm ond opsyonèl, ansanm ak espesifikasyon teknik ak gid antretyen.
Preview Gid rapid pou osiloskop pòtab OWON HDS200 Seri a
This quick guide provides essential information for operating the OWON HDS200 Series Handheld Oscilloscope. It covers product certification, safety precautions, detailed instructions for using the oscilloscope and multimeter functions, waveform generator operation, troubleshooting common issues, and maintenance procedures.
Preview OWON Digital Multimeter User Manual: B35(T)(+), D35(T), B41T(+)
Comprehensive user manual for OWON Digital Multimeters, models B35(T)(+), D35(T), and B41T(+). Covers safety, operation, features, technical specifications, and connectivity. Visit www.owon.com.cn for more information.
Preview Manyèl Itilizatè Osiloskop ak Miltimèt Depo Dijital Pòtab OWON HDS-N Seri
Manyèl itilizatè konplè pou Osiloskop ak Miltimèt Depo Dijital Pòtab OWON HDS-N Seri a, ki detaye operasyon, karakteristik, sekirite, ak espesifikasyon pou modèl HDS1021M-N, HDS2061M-N, ak HDS3101M-N.
Preview Manyèl Itilizatè Osiloskop OWON HDS100 Seri | Sekirite, Operasyon, ak Espesifikasyon
Manyèl itilizatè konplè pou mèt osiloskop seri OWON HDS100 la, ki kouvri enfòmasyon sekirite, gid demaraj rapid, operasyon miltimèt ak osiloskop, espesifikasyon teknik, ak antretyen. Aprann kijan pou itilize aparèy OWON ou a efektivman epi san danje.
Preview OWON HDS-Reihe LCD-Oszilloskop Bedienungsanleitung
Umfassende Bedienungsanleitung für das LCD-Oszilloskop OWON HDS-Reihe. Enthält Informationen zu Funktionen, Bedienung, technischen Daten und Fehlerbehebung für Modelle wie HDS242, HDS2102, HDS242S, HDS272, HDS272S, HDS2102S.