Joerns Healthcare HPL700

Joerns Hoyer HPL700 Bariatric Electric Patient Lift

Manyèl Enstriksyon Itilizatè

1. Entwodiksyon

This manual provides essential instructions for the safe and effective operation, maintenance, and troubleshooting of the Joerns Hoyer HPL700 Bariatric Electric Patient Lift. Designed for bariatric patients, this lift offers a robust 700 lbs (318 kg) weight capacity, a powered base, a 6-point bari loop cradle, and an integrated digital scale for precise patient handling.

Enfòmasyon enpòtan sou sekirite

Always read and understand all instructions and warnings before operating the patient lift. Failure to do so may result in serious injury or equipment damage.

  • Ensure the lift is on a stable, level surface before use.
  • Never exceed the maximum weight capacity of 700 lbs (318 kg).
  • Always use appropriate slings that are compatible with the 6-point bari loop cradle.
  • Regularly inspect the lift for any signs of wear, damage, or loose components.
  • Kenbe men ak rad lwen pati k ap deplase yo pandan w ap opere.
  • In case of emergency, utilize the emergency stop button and manual lowering functions.

2. Konpozan souview

The Joerns Hoyer HPL700 is comprised of several key components designed for safe and efficient patient transfer. Familiarize yourself with each part before operation.

Joerns Hoyer HPL700 Bariatric Electric Patient Lift

Yon plen view of the Joerns Hoyer HPL700 Bariatric Electric Patient Lift, showcasing its robust frame, adjustable boom, and wide base.

Konpozan kle:

  • Lift Mast and Boom: The vertical support structure and the adjustable arm that raises and lowers the patient.
  • 6-Point Bari Loop Cradle: The attachment point for slings, designed for secure and versatile patient positioning.
  • Digital Scale: Integrated weighing system for accurate patient weight measurement.
  • Powered Base: Electrically adjustable legs for easy access around furniture and obstacles.
  • Control Panel / Hand Control: Intuitive controls for operating the lift functions.
  • Hoyer Smart Monitor: Provides lift cycle count, overload protection, service reminders, and emergency functions.
  • Bouton pou kanpe ijans: Immediately cuts power to the lift in an emergency.
  • Manual Lowering Mechanism: Allows for manual lowering of the patient in case of power failure.
  • Woulèt: Wheels for mobility, typically with locking mechanisms.

3. Enstalasyon ak Premye Asanblaj

The Joerns Hoyer HPL700 typically arrives partially assembled. Follow these steps for final assembly and initial preparation.

  1. Debake: Retire tout konpozan yo nan anbalaj la avèk prekosyon. Enspekte pou wè si gen nenpòt domaj pandan transpò a.
  2. Asanble baz: Ensure the mast is securely attached to the base according to the included assembly diagrams. Verify all bolts are tightened.
  3. Enstalasyon batri ak chaje: Install the battery pack into its designated compartment. Connect the lift to a power outlet using the provided charger. Allow the battery to fully charge before first use (typically 8-12 hours). The lift can be operated while charging, but optimal performance requires a full charge.
  4. Tcheke fonksyonèl: After charging, perform a full functional check without a patient. Test all lift up/down, base open/close, and emergency stop functions.

4. Operating the Joerns Hoyer HPL700

Patient Transfer Procedures

Always ensure two caregivers are present for bariatric patient transfers when possible. Communicate clearly with the patient throughout the process.

  1. Prepare the Patient: Position the appropriate sling under the patient according to the sling manufacturer's instructions.
  2. Mete asansè a nan pozisyon sa a: Roll the lift to the patient. Use the powered base controls to open or close the legs as needed to position the lift securely around the patient's bed, chair, or wheelchair. Lock the casters once positioned.
  3. Attach the Sling: Connect the sling loops securely to the 6-point bari loop cradle. Double-check all connections.
  4. Raise the Patient: Using the hand control, slowly raise the patient a few inches off the surface. Pause to ensure the sling is correctly positioned and the patient is comfortable and secure. Continue raising the patient to the desired height for transfer.
  5. Transfè: Unlock the casters and carefully maneuver the lift to the destination. Lock the casters once in position.
  6. Lower the Patient: Slowly lower the patient onto the new surface. Once the patient is fully supported, unhook the sling from the cradle.
  7. Remove the Lift: Unlock the casters and carefully move the lift away from the patient.

Using the Digital Scale

Close-up of the digital scale display on the patient lift.

Yon detaye view of the integrated digital scale display, showing clear numerical readings for patient weight.

The integrated digital scale provides accurate weight measurements. To use:

  • Ensure the patient is fully suspended by the lift and not touching any other surface.
  • The scale display will automatically show the patient's weight.
  • Refer to the scale's specific instructions for changing units (pounds/kilograms) or taring the scale if a sling weight needs to be deducted.

Powered Base Operation

Close-up of the powered base mechanism of the patient lift.

Yon detaye view of the powered base, illustrating how the legs can be adjusted for access around furniture.

The powered base allows for easy adjustment of the lift's legs to navigate around obstacles.

  • Use the designated buttons on the hand control to open or close the base legs.
  • Open the legs for greater stability during transfers or to fit around wider furniture.
  • Close the legs for easier maneuvering through doorways or narrow spaces.

6-Point Bari Loop Cradle

Close-up of the 6-point bari loop cradle on the patient lift.

A close-up of the 6-point bari loop cradle, designed for secure and versatile sling attachment.

The 6-point bari loop cradle is designed for compatibility with various Hoyer slings, offering flexibility and comfort for different patient positions.

  • Ensure sling loops are correctly and securely attached to all six points of the cradle.
  • Always verify that the sling is properly seated and balanced before lifting.

Hoyer Smart Monitor

Close-up of the Hoyer Smart Monitor control panel.

Yon detaye view of the Hoyer Smart Monitor, displaying indicators for lift cycle, overload protection, service reminders, and emergency functions.

The Hoyer Smart Monitor provides critical operational data and safety features:

  • Lift Cycle Counter: Tracks the number of lift cycles for maintenance scheduling.
  • Pwoteksyon Surcharge: Prevents operation if the weight limit is exceeded.
  • Rapèl sèvis: Endike kilè antretyen routin dwe fèt.
  • Fonksyon Ijans: Provides visual and/or audible alerts for critical situations.

5. Antretyen ak Swen

Regular maintenance ensures the longevity and safe operation of your Joerns Hoyer HPL700.

  • Netwayaj chak jou: Wipe down all surfaces with a mild, non-abrasive cleaner and a damp cloth. Avoid harsh chemicals or excessive moisture.
  • Weekly Inspection: Check all bolts, nuts, and fasteners for tightness. Inspect the sling attachment points, casters, and electrical cables for any signs of wear or damage.
  • Swen batri: Keep the battery charged. Do not allow the battery to fully discharge frequently, as this can reduce its lifespan. Replace batteries when they no longer hold a sufficient charge.
  • Sèvis Pwofesyonèl: Adhere to the service reminders from the Hoyer Smart Monitor. Professional servicing should be performed by qualified technicians at recommended intervals.

6. Gid Depanaj

This section addresses common issues you might encounter with your patient lift.

PwoblèmKòz posibSolisyon
Lift pa opereBattery discharged; Emergency stop engaged; Power cable disconnected; Overload protection activated.Charge battery; Release emergency stop; Check power connections; Reduce load below 700 lbs.
Lift operates slowly or weaklyLow battery charge; Motor wear.Recharge battery fully; Contact service for motor inspection.
Base legs do not open/closeObstriksyon; Pwoblèm motè.Check for obstructions; Contact service if issue persists.
Digital scale shows incorrect readingPatient touching other surfaces; Calibration needed.Ensure patient is fully suspended; Refer to scale manual for calibration or contact service.
Bwi etranj pandan operasyon anLoose components; Motor issue; Lack of lubrication.Inspect for loose parts and tighten; Contact service for inspection and lubrication.

7. Espesifikasyon teknik

Detailed specifications for the Joerns Hoyer HPL700 Bariatric Electric Patient Lift.

Table of specifications for the HPL700 patient lift.

A table presenting detailed measurements and weights for the HPL700 patient lift, including overall length, height, leg widths, and component weights.

SpesifikasyonMeasurement (Imperial)Measurement (Metric)
Maximum Overall Length56"1423 mm
Minimum Overall Length55"1397 mm
Maksimòm Wotè An jeneral81"2058 mm
Maximum Overall Height (Boom)57.75"1467 mm
Legs Open - External Width46.25"1175 mm
Legs Open - Internal Width42.25"1073 mm
Legs Closed - External Width27.5"699 mm
Legs Closed - Internal Width23.5"597 mm
Limit Pwa700 liv (318 kg)
Pwa atik150 liv (68 kg)
MateryèlAlloy Steel
Kalite ekspozisyonLED with smart monitor
ManifaktiJoerns Swen Sante

8. Garanti ak sipò

Enfòmasyon sou Garanti

The Joerns Hoyer HPL700 comes with a limited lifetime warranty on the Hoyer frame and a 2-year warranty on the lift components. This warranty covers manufacturing defects under normal use. Please retain your proof of purchase for warranty claims.

Sipò pou Kliyan

For technical assistance, troubleshooting not covered in this manual, or warranty service, please contact Joerns Healthcare customer support. Refer to the official Joerns Healthcare websit la oswa dokiman acha ou a pou jwenn enfòmasyon kontak aktyèl.

Ou ka vizite tou Joerns Healthcare Store on Amazon pou plis enfòmasyon ak resous pwodwi yo.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.